Laser-treated stainless steel mini-screw implants: 3D surface roughness, bone-implant contact, and fracture resistance analysis

Date
2016-04
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Oxford University Press
Abstract

BACKGROUND/OBJECTIVES: This study investigated the biomechanical properties and bone-implant intersurface response of machined and laser surface-treated stainless steel (SS) mini-screw implants (MSIs). MATERIAL AND METHODS: Forty-eight 1.3mm in diameter and 6mm long SS MSIs were divided into two groups. The control (machined surface) group received no surface treatment; the laser-treated group received Nd-YAG laser surface treatment. Half in each group was used for examining surface roughness (Sa and Sq), surface texture, and facture resistance. The remaining MSIs were placed in the maxilla of six skeletally mature male beagle dogs in a randomized split-mouth design. A pair with the same surface treatment was placed on the same side and immediately loaded with 200 g nickel-titanium coil springs for 8 weeks. After killing, the bone-implant contact (BIC) for each MSI was calculated using micro computed tomography. Analysis of variance model and two-sample t test were used for statistical analysis with a significance level of P <0.05. RESULTS: The mean values of Sa and Sq were significantly higher in the laser-treated group compared with the machined group (P <0.05). There were no significant differences in fracture resistance and BIC between the two groups. LIMITATION: animal study CONCLUSIONS/IMPLICATIONS: Laser treatment increased surface roughness without compromising fracture resistance. Despite increasing surface roughness, laser treatment did not improve BIC. Overall, it appears that medical grade SS has the potential to be substituted for titanium alloy MSIs.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Kang, H.-K., Chu, T.-M., Dechow, P., Stewart, K., Kyung, H.-M., & Liu, S. S.-Y. (2016). Laser-treated stainless steel mini-screw implants: 3D surface roughness, bone-implant contact, and fracture resistance analysis. The European Journal of Orthodontics, 38(2), 154–162. http://doi.org/10.1093/ejo/cjv017
ISSN
1460-2210
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
European Journal of Orthodontics
Rights
Publisher Policy
Source
PMC